This page shows a
HTML5
canvas implementation of the patient support system. The table of
the system can move up/down and in/out to position patients in the bore of the MRI scanner.
Basic control requirements are:
- The table may only move in the bore when fully up
- The table may only move down when fully out
- The initialization may only start when the start button is pressed
and the tumble switch is in neutral
- When the tabletop release (TTR) switch is pressed, the table must stop
moving in the vertical direction and the table may move in horizontal
direction in an uncontrolled manner
- After the TTR switch is switched off, the system should re-initialize
More information on the system can be found in
this
document, in chapter 6.